The velocity factor of the receiving port can be defined using [Offset Embed] > "Offset"
(see "Permittivity / Velocity Factor" on page 543).
Remote command:
CALCulate<Chn>:TRANsform:TIME:XAXis
Rise Time/Rise Time Def.
Allows you to get and set the effective rise time for the configured time domain stimu-
lus.
This parameter is only available for low pass transformation types. It is inversely rela-
ted to the stop frequency.
Remote command:
[SENSe<Ch>:]HARMonic:RTIMe:THReshold
[SENSe<Ch>:]HARMonic:RTIMe:DATA
10.3 Scale softtool
The "Scale" softtool allows you to define how the active trace is displayed in its current
format.
Access: [Scale] hardkey
10.3.1 Scale Values tab
Provides the functions for diagram scaling.
The "Scale Values" settings are closely related to the "Format" and "Display" settings.
Scale softtool