Command reference
R&S
®
ZNL/ZNLE
947User Manual 1178.5966.02 ─ 20
Suffix:
<Ch>
.
Channel number
Parameters:
<DUTRiseTime> Default unit: s
Options: R&S ZNL/ZNLE-K2
Manual operation: See "Rise Time/Rise Time Def." on page 316
[SENSe<Ch>:]HARMonic:RTIMe:THReshold <DUTRiseThreshold>
Defines how the rise time that is set using [SENSe<Ch>:]HARMonic:RTIMe:DATA
shall be interpreted. Allows to select between rise time definitions 20%-80% (default)
and 10%-90%.
The set command automatically activates the "Automatic Harmonic Grid" function, i.e.
it sets [SENSe<Ch>:]HARMonic:AUTO to ON.
Suffix:
<Ch>
.
Channel number
Parameters:
<DUTRiseThreshold> T1_9 | T2_8
Rise time definition:
T1_9: 10%-90%
T2_8: 20%-80%
*RST: T2_8
Options: R&S ZNL/ZNLE-K2
Manual operation: See "Rise Time/Rise Time Def." on page 316
[SENSe:]LPORt...
The [SENSe:]LPORt... commands define the reference impedances of the bal-
anced ports.
[SENSe<Ch>:]LPORt<LogPt>:ZCOMmon........................................................................ 947
[SENSe<Ch>:]LPORt<LogPt>:ZDIFferent.........................................................................947
[SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe]...............................................................948
[SENSe<Ch>:]LPORt<LogPt>:ZCOMmon <RealPart>[, <ImaginaryPart>]
[SENSe<Ch>:]LPORt<LogPt>:ZDIFferent <RealPart>[, <ImaginaryPart>]
These commands specify the complex common mode and differential mode reference
impedances for the balanced (logical) port numbered <LogPt>.
Use [SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe] to toggle between config-
ured and default reference impedances.