EasyManua.ls Logo

R&S ZNL Series - Page 948

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNL/ZNLE
948User Manual 1178.5966.02 ─ 20
<LogPt> Logical port number. The logical ports must be defined using
SOURce<Ch>:LPORt<LogPt> <PhysicalPort1>,
<PhysicalPort2>. An n port analyzer supports a maximum of
n/2 (n even) or (n - 1)/2 (n odd) logical ports.
Parameters:
<RealPart> Real part of the port impedance.
Range: 1 mΩ to 10 MΩ
*RST: ZCOMmon: 25 Ω; ZDIFferent: 100 Ω (real impedan-
ces)
Default unit: Ohm
<ImaginaryPart> Imaginary part of the port impedance; may be omitted to define
a real impedance.
Range: - 10 MΩ to 10 MΩ
*RST: 0 Ω (real impedances)
Default unit: Ohm
Example:
*RST; :CALC:PAR:DEL 'TRC1'
Reset the analyzer and delete the (default) trace for channel no.
1.
SOUR:LPOR1 1,2
Combine the physical ports no. 1 and 2 to define the balanced
(logical) port no. 1.
LPOR1:ZDIF 27, 2
Specify a complex differential mode reference impedance of
27Ω + j * 2Ω for the defined port.
Manual operation: See "Single Ended Mode / Common Mode / Differential Mode"
on page 401
[SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe] <arg0>
Allows to toggle between default and renormalized reference impedance(s) for logical
port <LogPt>.
Suffix:
<Ch>
.
Channel number
<LogPt> Logical port number
Parameters:
<arg0> ON (1): Use default impedance(s)
OFF (0): Use the redefined impedances defined via
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence for single-ended
or via [SENSe<Ch>:]LPORt<LogPt>:ZCOMmon and
[SENSe<Ch>:]LPORt<LogPt>:ZDIFferent for balanced
logical ports
Manual operation: See "Use Default" on page 402
VNA command reference

Table of Contents

Related product manuals