Command reference
R&S
®
ZNL/ZNLE
700User Manual 1178.5966.02 ─ 20
CALCulate<Chn>:MARKer[:STATe]:AREA <HorizontalPos>, <VerticalPos>
Moves the marker info field for the active trace <Chn> to one of nine predefined posi-
tions in the active diagram.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos> LEFT | MID | RIGHt
Horizontal position
<VerticalPos> TOP | MID | BOTTom
Vertical position
Example:  See CALCulate<Chn>:MARKer<Mk>:Y
Manual operation:  See "Mkr 1 ... Mkr 10" on page 451
CALCulate<Chn>:MARKer<Mk>:TARGet <TargetSearchVal>
CALCulate<Chn>:MARKer<Mk>:REFerence:TARGet <TargetSearchVal>
Defines the target value for the target search of marker the related marker (see
"Marker addressing" on page 681).
The target search can be activated using CALCulate<Chn>:MARKer<Mk>:
FUNCtion:EXECute TARGet or CALCulate<Chn>:MARKer<Mk>:REFerence:
FUNCtion:EXECute TARGet.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number (ignored for reference markers)
Parameters:
<TargetSearchVal> Target search value of marker no. <Mk>.
The value range and reset value depend on the selected target
format (see CALCulate<Chn>:MARKer<Mk>:SEARch:
FORMat or CALCulate<Chn>:MARKer<Mk>:REFerence:
SEARch:FORMat).
Example: 
CALC:MARK ON
Create marker no. 1 and display it in the center of the sweep
range.
:CALC:MARK:TARG -10; FUNC:EXEC TARG
Define a target search value of -10 dB and start the target
search.
CALC:MARK:X?
Query the stimulus value corresponding to the target search
result.
Manual operation:  See "Target Value" on page 466
VNA command reference