Command reference
R&S
®
ZNL/ZNLE
669User Manual 1178.5966.02 ─ 20
Range: 0, 1, 2 (see above)
<StartStim>,
<StopStim>,
<StartResp>,
<StopResp>
Stimulus and response values of the first and last points of the
limit line segment.
The unit of the stimulus values is adjusted to the sweep type of
the active channel ([SENSe<Ch>:]SWEep:TYPE), the unit of
the ripple limit is adjusted to the format of the active trace
(CALCulate<Chn>:FORMat).
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 1.5 GHZ
Define an upper limit line segment in the stimulus range
between 1 GHz and 1.5 GHz, using default response values.
CALC:LIM:DATA 1, 1500000000, 2000000000, 2, 3
Define an upper limit line segment in the stimulus range
between 1.5 GHz and 2 GHz, assigning response values of +2
dB and +3 dB.
CALC:LIM:DISP ON
Show the limit line segment in the active diagram.
Manual operation: See "Segment List" on page 436
CALCulate<Chn>:LIMit:DCIRcle:CLEar
Resets the display circle to its default configuration (unit circle; show border: off; limit to
circle: off).
Suffix:
<Chn>
.
Channel number used to identify the active trace
Usage: Event
Manual operation: See "Clear Circle" on page 448
CALCulate<Chn>:LIMit:DCIRcle:DATA <CenterX>, <CenterY>, <Radius>
Defines the display circle for the active trace of channel <Chn>.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<CenterX> X position (real part) of the display circle's center
<CenterY> Y position (imaginary part) of the display circle's center
<Radius> Radius of the display circle
Manual operation: See "Draw Circle / Radius, Center X, Center Y" on page 448
CALCulate<Chn>:LIMit:DCIRcle:DISPlay[:STATe] <Boolean>
Sets/queries the visibility of the display circle for the active trace of channel <Chn>.
The display circle is defined using CALCulate<Chn>:LIMit:DCIRcle:DATA.
VNA command reference