Command reference
R&S
®
ZNL/ZNLE
678User Manual 1178.5966.02 ─ 20
Manual operation: See "Linear/Formula" on page 439
CALCulate<Chn>:LIMit:SEGMent<Seg>:INTerpol <LimLineInterpolation>
"Interpolation" determines whether the limit line segment is interpolated linearly or log-
arithmically. The latter may be more suitable for logarithmic sweeps.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Segment number
Parameters:
<LimLineInterpolation>LOGarithmic | LINear
Manual operation: See "Segment List" on page 436
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STARt <FreqPowTime>
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STOP <StimVal>
These commands change the start and stop stimulus values (i.e. the smallest and the
largest stimulus values) of a limit line segment. A segment must be created first to
enable the commands (e.g CALCulate<Chn>:LIMit:DATA). See also Chap-
ter 9.4.1.1, "Rules for limit line definition", on page 230.
Tip: To define the stimulus values of several limit line segments with a single com-
mand, use CALCulate<Chn>:LIMit:CONTrol[:DATA].
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Segment number
Parameters:
<StimVal> Stimulus value confining the limit line segment.
If specified, the unit of a stimulus value must in accordance with
the sweep type of the active channel ([SENSe<Ch>:]SWEep:
TYPE). Default units are Hz for frequency sweeps, dBm for
power sweeps (R&S ZNL only), and s for time sweeps. For CW
mode sweeps, stimulus values are dimensionless.
Default unit: NN
VNA command reference