General instrument setup
R&S
®
ZNL/ZNLE
138User Manual 1178.5966.02 ─ 20
Calibration Frequency MW
Uses the microwave calibration signal as the RF input. This function is used to cali-
brate the YIG-filter on the microwave converter. The microwave calibration signal is
pulsed. The microwave calibration signal frequency is fixed at 1.5 GHz.
Remote command:
DIAGnostic:SERVice:INPut[:SELect] on page 606
7.5.4 Service functions
Access: [Setup] > "Service" > "Service Function"
Using service functions
The service functions are not necessary for normal measurement operation. Incorrect
use can affect correct operation and/or data integrity of the R&S ZNL/ZNLE.
Therefore, many of the functions can only be used after entering a password. These
functions are described in the instrument service manual.
Service Function......................................................................................................... 138
Send ...........................................................................................................................139
Clear History............................................................................................................... 139
Password.................................................................................................................... 139
Clear Results...............................................................................................................139
Save Results...............................................................................................................139
Result list.....................................................................................................................139
Service Function
Selects the service function by its numeric code or textual name.
Service functions