Command reference
R&S
®
ZNL/ZNLE
658User Manual 1178.5966.02 ─ 20
●
Formatted trace data (FDATa) can only be imported to "live" traces if the related
channel is in single sweep mode (INITiate<Ch>:CONTinuous OFF).
Before the import, the target trace must be prepared according to the settings used
during export. Any request for new data from the hardware ("Restart Sweep" in sin-
gle sweep mode or switching to continuous sweep mode) discards imported data
and switches back the display to measured data.
Parameters:
<TraceName> String parameter containing the trace name
<Format> FDATa | SDATa | MDATa | NCData | UCData
Data format; see Table 11-5.
<Data> Trace data for FDATa | SDATa import either in ASCII or block
data format, depending on the current FORMat[:DATA] setting.
The column order must match the one used during export.
Example: See CALCulate<Chn>:DATA.
CALCulate:FILTer[:GATE]...
The CALCulate:FILTer[:GATE]... commands define the properties of the time
gate which is used to optimize the time domain response.
CALCulate:FILTer[:GATE]:TIME:AOFFset.........................................................................658
CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer.................................................................659
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev......................................................... 659
CALCulate<Chn>:FILTer[:GATE]:TIME:LINK..................................................................... 660
CALCulate<Chn>:FILTer[:GATE]:TIME:SHAPe..................................................................660
CALCulate<Chn>:FILTer[:GATE]:TIME:SHOW.................................................................. 661
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN....................................................................661
CALCulate<Chn>:FILTer[:GATE]:TIME:STARt...................................................................661
CALCulate<Chn>:FILTer[:GATE]:TIME:STOP................................................................... 661
CALCulate<Chn>:FILTer[:GATE]:TIME:STATe...................................................................662
CALCulate<Chn>:FILTer[:GATE]:TIME[:TYPE]..................................................................662
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow............................................................... 663
CALCulate:FILTer[:GATE]:TIME:AOFFset <Boolean>
Activates the operating mode where the time gate is moved in the opposite direction
when the "Delay" setting is changed.
Parameters:
<Boolean> ON | OFF - enable or disable "Adjust Time Gate".
*RST: OFF
VNA command reference