Command reference
R&S
®
ZNL/ZNLE
660User Manual 1178.5966.02 ─ 20
Parameters:
<SidebandSupp> Sideband suppression
Range:  10 dB to 120 dB
Increment:  10 dB
*RST:  32 dB
Default unit: dB
Example: 
*RST; :CALC:FILT:TIME:WIND DCH
Reset the instrument and select a Dolph-Chebyshev time gate
for filtering the data in the frequency domain.
CALC:FILT:TIME:DCH 25
Set the sideband suppression to 25 dB.
Manual operation:  See "Side Lobe Level" on page 367
CALCulate<Chn>:FILTer[:GATE]:TIME:LINK <TraceName>
Allows you to link the time gates of different (time domain) traces, i.e. to align the time
gate settings across linked (~coupled) traces.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TraceName> Name of the trace to be linked to the active trace of channel
<Chn>.
Use the empty string to unlink the active trace of channel
<Chn>.
Manual operation:  See "Time Gate Linking/Single Time Gate Linking" on page 367
CALCulate<Chn>:FILTer[:GATE]:TIME:SHAPe <TimeGate>
Selects the time gate to be applied to the time domain transform.
Tip:
Use the generalized command CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow
if you wish to select a Dolph-Chebychev time gate.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TimeGate> MAXimum | WIDE | NORMal | MINimum
MINimum - Steepest edges (rectangle)
WIDE - Normal gate (Hann)
NORM - Steep edges (Hamming)
Maximum - Maximum flatness (Bohman)
*RST:  WIDE
VNA command reference