Command reference
R&S
®
ZNL/ZNLE
1054User Manual 1178.5966.02 ─ 20
Diagrams are referenced by a window suffix <Wnd>. An additional suffix <WndTr>
in the DISPlay:WINDow<Wnd>:TRACe<WndTr>... commands numbers the dif-
ferent traces in a diagram.
In remote control, it is possible to display the same trace in several diagrams.
The analyzer provides several commands allowing a smooth transition between
remote and manual control.
// 1. One channel, two traces, one diagram
// Reset the instrument, creating the default trace Trc1 in channel 1.
// The default measured quantity is the forward transmission S-parameter S21.
// The default format is dB Mag.
*RST
// Create a second trace in channel 1, assign the format Phase,
// and display the new trace in the same diagram.
// the trace becomes the active trace but is not displayed
CALCulate1:PARameter:SDEFine 'Trc2', 'S21'
// the trace is referenced by the channel suffix 1
CALCulate1:FORMat PHASe
// display the second trace, numbering it the second trace in diagram no. 1
DISPlay:WINDow1:TRACe2:FEED 'Trc2'
// Check the result on the local screen
// Go to local
SYSTem:DISPlay:UPDate ONCE
// 2. One channel, two traces, two diagrams
// Create a second diagram, assign Trc2 to the new area, and remove it
// from the first area.
DISPlay:WINDow2:STATe ON
DISPlay:WINDow2:TRACe2:FEED 'Trc2'
// Trc2 is now displayed in both diagrams
DISPlay:WINDow1:TRACe2:DELete
// Check the result on the local screen
// Go to local
SYSTem:DISPlay:UPDate ONCE
// 3. One channel, four traces, four diagrams
// Reset the instrument, add diagrams no. 2, 3, 4.
*RST; :DISPlay:WINDow2:STATe ON
DISPlay:WINDow3:STATe ON
DISPlay:WINDow4:STATe ON
VNA programming examples