EasyManua.ls Logo

R&S ZNL Series - Page 191

R&S ZNL Series
1135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VNA concepts and features
R&S
®
ZNL/ZNLE
191User Manual 1178.5966.02 ─ 20
Note that the channel data flow for S-parameters (and quantities derived from S-
parameters such as impedances, admittances, stability factors) differs from the chan-
nel data flow for wave quantities (and derived quantities such as ratios).
RATIOS
Differential
Match
EMBEDDING
IMBALANCE
IMPEDANCE
RENORMA-
LIZATION
OFFSET
(Alternative 2)
EMBEDDING
Ground Loop >
Balanced >
Port Set >
Single Ended
DEEMBEDDING
Single Ended >
Port Set >
Balanced >
Ground Loop
SYSTEM
ERROR
CORRECTION
(User)
POWER
CORRECTION
(User)
R&S ZNL only
RAW WAVE
QUANTITIES
SYST. ERR.
CORRECTION
DATA
POWER
CORRECTION
DATA
OFFSET
(Alternative 1)
MIXED MODE
S-PARAMS
AVERAGE
COMPLEX
CONVERSION
**
S-parameters
Wave quantities with
wave de-/embedding
Wave quantities without
wave de-/embedding
SYSTEM
ERROR
CORRECTION
(Factory)
0
a b
1
2
3
Data access points:
0 Raw wave quantities
1 Factory-calibrated S-parameters
2 S-parameters
3 Balanced (symmetric) S-parameters
a, b apply cal pool data /
copy to cal pool
– S-parameters (no conversion)
– CMRR
– Converted impedances
– Converted admittances
– Stability factors
**
Traces...
Channel data flow
for all traces
of the channel
Trace data flow
for all traces of the channel
Figure 9-1: Channel data flow
Basic concepts

Table of Contents

Related product manuals