What's new in firmware version 1.71
R&S
®
ZNL/ZNLE
22User Manual 1178.5966.02 ─ 20
"Max Hold" traces: "Restart" and "Restart All" did not work
A memory trace with "Math" could not be recalled correctly if the original trace was
a wave
Considerable residual responses for some configurations
The analyzer GUI became slow if many markers were active
In FW versions 1.61 and higher, selecting a trace via
CALCulate<Ch>:PARameter:SELect <TraceName> restarted the sweep
Spectrum analysis changes
The spectrum analysis function of the R&S ZNL (with hardware option B1) and its addi-
tional software options are based on firmware 2.00 of the Rohde & Schwarz spectrum
analyzer R&S FPL1000.
See the Release Notes of the R&S FPL1000 (at https://www.rohde-schwarz.com/firm-
ware/fpl1000/) for a history of changes.
Firmware version 1.71