VNA GUI reference
R&S
®
ZNL/ZNLE
330User Manual 1178.5966.02 ─ 20
Tip: You can change to point based x-axis to improve the display of a segmented fre-
quency sweep (see "Seg X-Axis" on page 332).
Remote command:
[SENSe<Ch>:]SWEep:TYPE SEGMent
Power
(R&S ZNL only!)
In a "Power" sweep, the internal generator power is swept in dB-linear, equidistant
steps over a continuous power range. The generator power range (sweep range) and
the fixed frequency can be specified in the "Stimulus" settings (see Chapter 10.2.1,
"Stimulus tab", on page 311).
"Power" sweeps are particularly suitable for the analysis of non-linear effects (satura-
tion, compression) on active and passive DUTs (e.g. power amplifiers, mixers).
In a Cartesian diagram, the measurement result is displayed as a trace over a dB-lin-
ear power scale. The following example shows a "Power" sweep in the source power
range between –25 dBm and 0 dBm, performed at a CW frequency of 1 GHz.
Any generator power calibration is included.
Remote command:
[SENSe<Ch>:]SWEep:TYPE POWer
Sweep Softtool