VNA GUI reference
R&S
®
ZNL/ZNLE
340User Manual 1178.5966.02 ─ 20
In particular, it is possible to use two different external trigger sources (R&S ZNL
only).
Remote command:
TRIGger<Ch>[:SEQuence]:SOURce EXTernal
Manual Trigger
Generates the trigger event for "Manual" trigger mode and is disabled unless this mode
is active.
Remote command:
TRIGger<Ch>[:SEQuence]:SOURce MANual
*TRG
Sequence
Selects the measurement cycle or sequence of actions to be triggered in "External" or
"Manual" mode.
●
"Sweep" means that each trigger event starts an entire sweep, according to the
current sweep configuration.
●
"Point" means that each trigger event starts the measurement at the next sweep
point.
●
"Partial Measurement" means that each trigger event starts the next partial mea-
surement at the current or at the next sweep point. If every sweep point only
requires a single partial measurement, this option is equivalent to "Point".
See also Chapter 9.1.4.1, "Partial measurements and driving mode", on page 188.
●
"Segment" means that each trigger event starts the next sweep segment within the
current sweep. If a sweep type other than Segmented is active, this option is equiv-
alent to "Sweep".
Trigger sequence and driving mode
Sweep Softtool