10.11.1.1 Controls on the Peak tab
Max / Min
Sets the active marker to the absolute maximum or minimum in the search range, i.e.
to the largest or smallest of all response values. If a complex trace format (e.g. a polar
diagram) is active, the marker is set to the measurement point with the maximum or
minimum magnitude.
"Max" and "Min" also overwrite the current "Search Mode" (--> "Search Min" and
"Search Max") and the "Peak Type" for the peak search functions.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MINimum | MAXimum
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:EXECute MINimum |
MAXimum
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:RESult?
Center = Marker
See "Center = Marker / Start = Marker / Stop = Marker / Span = Marker" on page 455.
Next Peak
Sets the active marker to the next local maximum or minimum in the search range,
depending on the selected Peak Type.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute NPEak
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:EXECute NPEak
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
CALCulate<Chn>:MARKer<Mk>:REFerence:FUNCtion:RESult?