Due to the R&S ZNL/ZNLE's calibration wizard, calibration is a straightforward, guided
process.
1. Replace the DUT by the Through standard of your calibration kit. Make sure to dis-
connect all calibration units.
2. Select [Cal] > "Start... (Manual)" to open the "Calibration Setting" wizard.
3. Select the port combination Port 1 (P1) and Port 2 (P2) and the calibration type
"Trans Norm". Make sure to define port 1 as the source port.
4. Select "Next" to proceed to the next page of the "Calibration Setting" wizard.
5. Select the test port connector type and gender (here: N 50 Ω, female, correspond-
ing to a male Through standard), and the calibration kit (here: R&S ZV-Z121).
6. Tap"Start".