Command reference
R&S
®
ZNL/ZNLE
626User Manual 1178.5966.02 ─ 20
The active traces are handled as follows:
After a preset (*RST), the analyzer displays a single diagram area with the default
trace no. 1 named Trc1. The trace is active in manual and in remote control.
In manual control, a new, added trace automatically becomes the active trace. To
select another trace as the active trace, tap inside the trace list.
In remote control, a new trace added via CALCulate<Ch>:PARameter:SDEFine
'<trace_name>', '<parameter>' also becomes the active trace. To select
another trace as the active trace, use (CALCulate<Ch>:PARameter:SELect
'<trace_name>').
The active traces for manual and remote control may be different.
Example:
*RST
Reset the analyzer, creating channel no. 1 with the default trace "Trc1". The trace is
displayed in diagram area no. 1.
CALC1:PAR:SDEF 'Trc2', 'S11'; DISP:WIND:TRAC2:FEED 'Trc2'
Create a new trace named "Trc2", assigned to channel no. 1 (the suffix 1 after CALC,
may be omitted), and display the trace. The new trace automatically becomes the
active trace for manual and for remote control. To check this, tap "Trace – Marker –
Marker 1" to create a marker. The marker is assigned to "Trc2". Delete all markers
("Trace – Marker – All Markers Off").
CALC1:MARK ON
Example:
To verify that "Trc2" is also active for remote control, use the channel suffix 1 after
CALC (may be omitted) to reference the active trace in channel 1 and create a marker
"Mkr 1". The marker is assigned to "Trc2".
Example:
CALC:PAR:SEL 'Trc1'; CALC1:MARK ON
Select the old default trace "Trc1" as the active trace for remote control. Create a new
marker to verify that "Trc1" is now the active trace in channel 1.
In the SCPI command description, the numeric suffix <Ch> is used for channel settings
(it denotes the configured channel), whereas <Chn> is used for trace settings (it refers
to the active trace in the channel).
11.4.2.3 Initiating measurements, speed considerations
After a reset the network analyzer measures in continuous mode. The displayed trace
shows the result of the last sweep and is continuously updated. This provides a perma-
nent visual control over the measurement and the effect of any analyzer settings.
In remote control, it is advisable to follow a different approach in order use the analyz-
er's resources to full capacity and gain measurement speed. The following principles
VNA remote control basics