Getting Started
R&S
®
ZNL/ZNLE
66User Manual 1178.5966.02 ─ 20
To optimize the display
To optimize the display for the calibration signal, we will adjust the main measurement
settings.
1. Set the center frequency to the calibration frequency:
a) Tap the "Overview" softkey to display the configuration "Overview".
b) Tap the "Frequency" button.
c) In the "Center" field, enter 50 on the number pad on the front panel.
d) Press the "MHz" key next to the number pad.
2. Reduce the span to 20 MHz:
a) In the "Span" field of the "Frequency" dialog box, enter 20
 
MHz.
b) Close the "Frequency" dialog box.
3. Set the reference level to -25 dBm:
a) In the configuration "Overview", tap the "Amplitude" button.
b) In the "Value" field of the "Amplitude" dialog box, enter -25
 
dBm.
The display of the calibration signal is now improved. The maximum at the center
frequency (=calibration frequency) of 50 MHz becomes visible.
Figure 4-12: Calibration signal with optimized display settings
Trying out the instrument