Command reference
R&S
®
ZNL/ZNLE
795User Manual 1178.5966.02 ─ 20
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
TNDefinition <CircuitModel>
Selects the circuit model for single ended port embedding.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<CircuitModel> FIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL |
SHLC | GSL | LSG | GSG | SLG | SGL | SGG
Possible circuit models (character data); see Table 11-9
*RST: CSL
Example: See CALCulate<Ch>:TRANsform:VNETworks:SENDed:
EMBedding<PhyPt>[:STATe]
Manual operation: See "Network" on page 551
CALCulate:TRANsform... (other)
CALCulate<Chn>:TRANsform:COMPlex <Result>
Converts S-parameters into converted (matched-circuit) Y-parameters or Z-parameters
and vice versa, assuming that port no. i is terminated with Z
0i
so that the three parame-
ter sets are equivalent and the following formulas apply:
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Result> S | Y | Z
S-parameters, Y-parameters, Z-parameters
VNA command reference