Command reference
R&S
®
ZNL/ZNLE
876User Manual 1178.5966.02 ─ 20
Pins 16 to 19 may be reserved for monitoring the drive ports 1 to 4 of the analyzer
(OUTPut<Ch>:UPORt:ECBits OFF) . This leaves up to 16 different monitored
channel states.
Tip: You can use the active channel number as a parameter for
OUTPut<Ch>:UPORt[:VALue] and monitor the activity of up to 255 different chan-
nels at the USER PORT connector; see example below. You can also use the USER
PORT output signals as channel-dependent (or drive port-dependent) trigger signals
for external devices. Furthermore you can use CONTrol:AUXiliary:C[:DATA] to
transfer the eight bit value in decimal representation.
Suffix:
<Ch>
.
Channel number
Parameters:
<BinValue> Binary value. The values correspond to the following states of
the USER PORT connector:
#B00000000 - no signal at any of the eight pins 8, 9, 10, 11, 16,
17, 18, 19
#B00000001 - output signal at pin 8
#B00000010 - output signal at pin 9
#B00000011 - output signal at pins 8 and 9
...
#B11111111 - output signal at pins 8, 9, 10, 11, 16, 17, 18, 19
Range: #B00000000 to #B11111111 (for setting command),
0 to 255 (query)
*RST: #B00000000 (0)
Example:
*RST; :OUTP1:UPOR #B00000001
Assign the channel bit value #B00000001 to the active channel
no. 1. The analyzer performs a measurement in channel no. 1,
therefore the output signal at pin 8 is switched on.
CONF:CHAN2:STAT ON; OUTP2:UPOR #B00000010
Create channel no. 2, causing it to become the active channel,
and assign the channel bit value #B00000010. The analyzer
performs no measurement in channel no. 2, therefore the output
signal is not changed.
CALC2:PAR:SDEF 'Ch2Tr1', 'S11'
Create a trace named 'Ch2Tr1' and assign it to channel 2. While
the analyzer measures in channel 2, the output signal changes
from pin 8 to pin 9.
OUTP:UPOR:ECB OFF
Reserve pin 16 to 19 for monitoring the drive ports of the ana-
lyzer.
Manual operation:
See "Channel Bits (Decimal)" on page 532
11.5.2.13 PROGram commands
The PROGram... commands control external application programs that can be run on
the analyzer.
VNA command reference

Table of Contents

Other manuals for R&S ZNL Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNL Series and is the answer not in the manual?

R&S ZNL Series Specifications

General IconGeneral
BrandR&S
ModelZNL Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals