R&S
®
ZNL/ZNLE
926User Manual 1178.5966.02 ─ 20
Calibration type Parameters in
[SENSe<Ch>:]CORRection:
COLLect:METHod:DEFine
Available error terms (depending on port num-
bers)
One-port normalization (reflection)
using an open or a short standard
REFL, RSHort 'REFLTRACK'
Full one port ("Refl OSM") FOPort 'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK'
Two-port normalization FRTRans 'TRANSTRACK'
One path two port OPTPort 'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'TRANSTRACK'
TOSM TOSM 'DIRECTIVITY', 'SRCMATCH', 'REFLTRACK',
'LOADMATCH', 'TRANSTRACK'
[SENSe<Ch>:]CORRection:CONNection <ConnectorName>[, <Mode>[, <Gender>[,
<RelPermittivity>[, <RefImpedance>]]]]
Configures the user-defined connector types.
Suffix:
<Ch>
.
Channel number
Parameters:
<ConnectorName> Name of the user-defined connectors, string parameter.
<Mode> TEM | WGUide
Transverse electric or waveguide type propagation mode.
<Gender> GENDer | NGENder
GENDer polar connector type (m/f)
NGENder sexless connector type
<RelPermittivity> Relative permittivity
Range: 0.0000000001 to 1000.
<RefImpedance> For TEM type connectors: reference impedance in Ω (without
unit), For waveguide (WGUide) type connectors: cutoff frequency
in Hz (without unit).
Range: Ref. impedance: 1µΩ to 1000 MΩ. Cutoff fre-
quency: 0 Hz to 1000 GHz.
*RST: -
Example:
CORR:CONN 'USERCON',TEM,GEND,1.00000,50
Define a TEM type connector type named USERCON.
CORR:CONN? 'USERCON'
Query the properties of the configured connector type.
CORR:CONN:DEL 'USERCON'
Delete the configured connector type.
Manual operation: See "Connector / Gender" on page 484