Remote Control
R&S
®
FSVA/FSV
1047Operating Manual 1176.7510.02 ─ 09
Result
Figure 4-10: Result: signal only (no gaps, no noise)
Example 2: Noise only
The following sample program records the signal in edge mode and displays only the
noise.
Used values
sample rate: 32 MHz
number of IQ samples: 128000
delay = 0 samples
length =9000 samples
trigger mode: EXT
Program
:TRACe1:IQ ON
Trace:iq:set NORM,0,32000000,EXT,POS, 0,128000
:TRACe1:IQ:EGATe ON
:TRACe1:IQ:EGATe:TYPE EDGE
:TRACe1:IQ:EGATe:LENGth 9000
Remote Control – Programming Examples