EasyManua.ls Logo

Rohde & Schwarz FSVA - Page 220

Rohde & Schwarz FSVA
1106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Functions
R&S
®
FSVA/FSV
220Operating Manual 1176.7510.02 ─ 09
Remote command:
FREQ:CENT:STEP:LINK SPAN, see [SENSe:]FREQuency:CENTer:STEP:LINK
on page 806
FREQ:CENT:STEP:LINK:FACT 50PCT, see [SENSe:]FREQuency:CENTer:STEP:
LINK:FACTor on page 806
0.5*RBW (span > 0) ← CF Stepsize
Sets the step size for the center frequency to 50 % of the resolution bandwidth.
Remote command:
FREQ:CENT:STEP:LINK RBW, see [SENSe:]FREQuency:CENTer:STEP:LINK
on page 806
FREQ:CENT:STEP:LINK:FACT 50PCT, see [SENSe:]FREQuency:CENTer:STEP:
LINK:FACTor on page 806
x*Span (span > 0) ← CF Stepsize
Opens an edit dialog box to set the step size for the center frequency as a percentage
(%) of the span.
Remote command:
FREQ:CENT:STEP:LINK SPAN, see [SENSe:]FREQuency:CENTer:STEP:LINK
on page 806
FREQ:CENT:STEP:LINK:FACT 20PCT, see [SENSe:]FREQuency:CENTer:STEP:
LINK on page 806
x*RBW (span > 0) ← CF Stepsize
Opens an edit dialog box to set the step size for the center frequency as a percentage
(%) of the resolution bandwidth. Values between 1 % and 100 % in steps of 1 % are
allowed. The default setting is 10 %.
Remote command:
FREQ:CENT:STEP:LINK RBW, see [SENSe:]FREQuency:CENTer:STEP:LINK
on page 806
FREQ:CENT:STEP:LINK:FACT 20PCT, see [SENSe:]FREQuency:CENTer:STEP:
LINK on page 806
=Center ← CF Stepsize
Sets the step size to the value of the center frequency and removes the coupling of the
step size to span or resolution bandwidth.
This function is especially useful for measurements of the signal harmonics. In this
case, each stroke of the arrow key selects the center frequency of another harmonic.
=Marker ← CF Stepsize
Sets the step size to the value of the current marker and removes the coupling of the
step size to span or resolution bandwidth.
This function is especially useful for measurements of the signal harmonics. In this
case, each stroke of the arrow key selects the center frequency of another harmonic.
Manual ← CF Stepsize
Opens an edit dialog box to enter a fixed step size for the center frequency.
Measurement Parameters

Table of Contents