Remote Control
R&S
®
FSVA/FSV
806Operating Manual 1176.7510.02 ─ 09
[SENSe:]FREQuency:CENTer:STEP:AUTO <State>
This command couples the step size of the center frequency to the span (ON) or sets
the value of the center frequency entered via [SENSe:]FREQuency:CENTer (OFF).
Parameters:
<State> ON | OFF
*RST: ON
Example:
FREQ:CENT:STEP:AUTO ON
Activates the coupling of the step size to the span.
[SENSe:]FREQuency:CENTer:STEP:LINK
<CouplingType>
This command couples and decouples the center frequency step size to the span or
the resolution bandwidth.
Parameters:
<CouplingType> SPAN
Couples the step size to the span. Available for measurements
in the frequency domain.
RBW
Couples the step size to the resolution bandwidth. Available for
measurements in the time domain.
OFF
Decouples the step size (manual input).
*RST: SPAN
Example:
FREQ:CENT:STEP:LINK SPAN
Manual operation: See "0.1*Span (span > 0)" on page 219
See "0.1*RBW (span > 0)" on page 219
See "0.5*Span (span > 0)" on page 219
See "0.5*RBW (span > 0)" on page 220
See "x*Span (span > 0)" on page 220
See "x*RBW (span > 0)" on page 220
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor <Factor>
This command defines a step size factor if the center frequency step size is coupled to
the span or the resolution bandwidth.
Parameters:
<Factor> Range: 1 to 100
*RST: 10
Default unit: PCT
Example:
FREQ:CENT:STEP:LINK:FACT 20PCT
Remote Control – Commands