Instrument Functions
R&S
®
FSVA/FSV
333Operating Manual 1176.7510.02 ─ 09
In adjacent-channel power measurement with one or several carrier signals, the power
is always referenced to a transmission channel, i.e. no value is displayed for "Tx1 (Ref)
Power".
Remote command:
[SENSe:]POWer:ACHannel:REFerence:AUTO ONCE on page 848
User Standard ← Ch Power ACLR
Opens a submenu to configure customized standards.
Load ← User Standard ← Ch Power ACLR
Opens a dialog to select and load a user defined ACLR standard.
Note: Compatibility to R&S FSP. User standards created on an analyzer of the R&S
FSP family are compatible to the R&S FSV. User standards created on an R&S FSV,
however, are not necessarily compatible to the analyzers of the R&S FSP family and
may not work there.
Remote command:
Querying available standards:
CALCulate<n>:MARKer<m>:FUNCtion:POWer:STANdard:CATalog?
on page 705
Loading a standard:
CALCulate<n>:MARKer<m>:FUNCtion:POWer:PRESet on page 699
Save ← User Standard ← Ch Power ACLR
Saves the current ACLR configuration in an xml file in order for you to use it again at a
later time. You can define the drive, path and file name in the corresponding dialog.
The default location is C:\R_S\Instr\acp_std\.
Note that the ACLR user standard is not supported by Fast ACLR and Multi Carrier
ACLR measurements.
If you create your own standard, you can customize the following parameters:
number of adjacent channels
channel bandwidth of transmission (Tx), adjacent (Adj) and alternate (Alt) channels
channel spacings
resolution and video bandwidth
ACLR limits and their state
sweep time and sweep time coupling
detector
trace mode
Remote command:
Configuring channels:
see "SENSe:POWer Subsystem" on page 841
Saving custom channel configurations:
CALCulate<n>:MARKer<m>:FUNCtion:POWer:STANdard:SAVE on page 705
Delete ← User Standard ← Ch Power ACLR
Deletes the user standard that you select in the corresponding dialog box. Note that
the R&S FSVA/FSV deletes the file without further notice.