Remote Control
R&S
®
FSVA/FSV
699Operating Manual 1176.7510.02 ─ 09
Parameters:
<Mode> WRITe | MAXHold
WRITe
The channel power and the adjacent channel powers are calcu-
lated directly from the current trace
MAXHold
The power values are calculated from the current trace and
compared with the previous power value using a maximum algo-
rithm.
Example:
CALC:MARK:FUNC:POW:MODE MAXH
Sets the Maxhold channel power mode.
Manual operation: See "Clear/Write" on page 331
See "Max Hold" on page 331
CALCulate<n>:MARKer<m>:FUNCtion:POWer:PRESet
<Standard> |
<UserStandard>
This command selects the power measurement setting for a standard and switches on
the corresponding measurement, if required.
The configuration for a standard comprises of the parameters weighting filter, channel
bandwidth and spacing, resolution and video bandwidth, as well as detector and sweep
time.
The settings for standards IS95A and C differ as far as the calculation method of chan-
nel spacings is concerned. For IS95A and J-STD008 the spacing is calculated from the
center of the main channel to the center of the corresponding adjacent channel, for
IS95C from the center of the main channel to the nearest border of the adjacent chan-
nel.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Setting parameters:
<Standard> Note that predefined standards go without quotes, while user
standards have to be in quotes ('user standard').
<Predefined standard>
Predefind standard according to table in "Predefined CP/ACLR
Standards" on page 711.
'<string>'
User-defined standard with <string> being the name of the user
standard.
Remote Control – Commands