Instrument Functions
R&S
®
FSVA/FSV
540Operating Manual 1176.7510.02 ─ 09
To open the Power Sensor menu
The power sensor option R&S FSV–K9 is integrated within the "Input/Output" menu.
Press the INPUT/OUTPUT key.
Press the "Power Sensor" softkey.The "Power Sensor" menu is displayed.
Menu and softkey description
Chapter 3.9.4, "Softkeys of the Power Sensor Menu (R&S FSV-K9)",
on page 545
Chapter 3.9.5, "Power Sensor Configuration Dialog", on page 547
Remote operation
Power sensors can be operated using the following remote command subsystems:
"CALCulate:PMETer Subsystem (Power Sensor, R&S FSV-K9)" on page 729
Chapter 4.2.3.3, "CALibration:PMETer Subsystem (Power Sensor, R&S FSV-K9)",
on page 744
Chapter 4.2.3.6, "FETCh:PMETer Subsystem (Power Sensor, R&S FSV-K9)",
on page 755
Chapter 4.2.3.14, "READ:PMETer Subsystem", on page 776
"SENSe:PMETer<p> Subsystem (Power Sensor, R&S FSV-K9)" on page 834
Chapter 4.2.3.21, "UNIT Subsystem", on page 911
Results display
The results of the power sensor measurements are displayed in the marker table. For
each power sensor, a row is inserted. The sensor index is indicated in the "Function"
column.
Alternatively, you can query the results using the remote command FETCh<n>:
PMETer<p>? on page 755.
Tasks
The following tasks with Power Sensors are described in this chapter:
Chapter 3.9.2, "Zeroing the Power Sensor", on page 542
Chapter 3.9.1, "Using Power Sensors", on page 541
Chapter 3.9.3, "Configuring an External Power Trigger", on page 543
Instrument Functions - Power Sensor (R&S