Remote Control
R&S
®
FSVA/FSV
616Operating Manual 1176.7510.02 ─ 09
Parameters:
<Mode> POWer
Result is displayed as a power in dBm.
DENSity
Result is displayed as a density in dBm/Hz.
*RST: POW
Example:
CALC:DELT:BPOW:MODE DENS
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?
This command queries the results of the band power measurement.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Return values:
<Power> Signal power over the delta marker bandwidth.
Example: Activate the delta band power marker:
CALC:DELT:BPOW:STAT ON
Select the density mode for the result:
CALC:DELT:BPOW:MODE DENS
Query the result:
CALC:DELT:BPOW:RES?
Response:
20dBm/Hz
Usage: Query only
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:SPAN <Span>
This command defines the bandwidth around the delta marker position. The span is
indicated by lines in the diagram.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<Span> Numeric value that defines the span in Hz.
The maximum span depends on the marker position and
R&S FSVA/FSV model.
*RST: 5% of current span
Example:
CALC:DELT:BPOW:SPAN 20
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer[:STATe] <State>
This command turns delta markers for band power measurements on and off.
Remote Control – Commands