Remote Control
R&S
®
FSVA/FSV
617Operating Manual 1176.7510.02 ─ 09
If neccessary, the command also turns on a reference marker.
For details see Chapter 3.3.3.8, "Performing Band Power Measurements",
on page 299.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<State> ON | OFF
When switched to on, if no marker is active yet, marker 1 is acti-
vated. Otherwise, the currently active marker is used as a delta
band power marker (all other marker functions for this marker
are deactivated).
*RST: OFF
Example:
CALC:DELT:BPOW:STAT ON
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed[:STATe] <State>
This command switches the relative measurement to a fixed reference value on or off.
Marker 1 is activated previously and a peak search is performed, if necessary. If
marker 1 is activated, its position becomes the reference point for the measurement.
The reference point can then be modified with the CALCulate<n>:
DELTamarker<m>:FUNCtion:FIXed:RPOint:X commands and CALCulate<n>:
DELTamarker<m>:FUNCtion:FIXed:RPOint:Y independently of the position of
marker 1 and of a trace. It applies to all delta markers as long as the function is active.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:DELT:FUNC:FIX ON
Switches on the measurement with fixed reference value for all
delta markers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm.
Manual operation: See "Ref. Fixed On/Off" on page 290
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
This command moves the fixed reference marker to the peak power.
Remote Control – Commands