Remote Control
R&S
®
FSVA/FSV
619Operating Manual 1176.7510.02 ─ 09
Parameters:
<RefPointLevel> Numeric value that defines the vertical position of the reference.
The unit and value range is variable.
*RST: Fixed reference: OFF
Example:
CALC:DELT:FUNC:FIX:RPO:Y -10dBm
Sets the reference point level for delta markers to -10 dBm.
Manual operation: See "Ref Point Level" on page 289
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y:OFFSet
<Offset>
This command defines a level offset for the fixed delta marker reference point.
When measuring the phase noise, the command defines an additional level offset for
delta marker 2.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<Offset> Level offset that is added to the display of all delta markers.
*RST: 0 dB
Example:
CALC:DELT:FUNC:FIX:RPO:Y:OFFS 10dB
Sets the level offset for the measurement with fixed reference
value or the phase-noise measurement to 10 dB.
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise[:STATe] <State>
This command turns the phase noise measurement at the delta marker position on and
off.
The correction values for the bandwidth and the log amplifier are taken into account in
the measurement.
The reference marker for phase noise measurements is either a normal marker or a
fixed reference. If necessary, the command turns on the reference marker
A fixed reference point can be modified with the CALCulate<n>:DELTamarker<m>:
FUNCtion:FIXed:RPOint:X and CALCulate<n>:DELTamarker<m>:FUNCtion:
FIXed:RPOint:Y commands independent of the position of marker 1 and of a trace.
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Note: marker 2 is always the deltamarker for phase noise mea-
surement results.
Parameters:
<State> ON | OFF
*RST: OFF
Remote Control – Commands