Remote Control
R&S
®
FSVA/FSV
620Operating Manual 1176.7510.02 ─ 09
Example:
CALC:DELT:FUNC:PNO ON
Switches on the phase-noise measurement with all delta mark-
ers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm
Manual operation: See "Phase Noise On/Off" on page 289
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:AUTO <State>
This command turns an automatic peak search for the fixed reference marker at the
end of a sweep on and off.
Suffix:
<n>
.
Selects the measurement window.
<m> irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:DELT:FUNC:PNO:AUTO ON
Activates an automatic peak search for the reference marker in
a phase-noise measurement.
Manual operation: See "Ph. Noise Auto Peak Search" on page 289
CALCulate<n>:DELTamarker<m>:FUNCtion:PNOise:RESult?
This command queries the result of the phase noise measurement.
If necessary, the command activates the measurement first..
Suffix:
<n>
.
Selects the measurement window.
<m> 2
Marker 2 is always the deltamarker for phase noise measure-
ments.
Example:
CALC:DELT2:FUNC:PNO:RES?
Outputs the result of phase-noise measurement of the delta-
marker 2.
Usage: Query only
Manual operation: See "Phase Noise On/Off" on page 289
CALCulate<n>:DELTamarker<m>:LINK <State>
This command links delta marker 1 to marker 1.
Remote Control – Commands