Remote Control
R&S
®
FSVA/FSV
650Operating Manual 1176.7510.02 ─ 09
Example:
CALC:LIM2:CONT 1 MHz,30 MHz,100 MHz,300 MHz,1
GHz
Defines 5 reference values for the x-axis of limit line 2.
CALC:LIM2:CONT?
Outputs the reference values for the x-axis of limit line 2 separa-
ted by a comma.
Manual operation: See "Edit Value" on page 418
CALCulate<n>:LIMit<k>:CONTrol:DOMain <SpanSetting>
This command selects the domain of the limit line.
Suffix:
<n>
.
Selects the measurement window.
<k> 1...8
Selects the limit line.
Parameters:
<SpanSetting> FREQuency | TIME
*RST: FREQuency
Example:
CALC:LIM2:CONT:DOM TIME
Defines zero span for the x-axis of limit line 2.
CALCulate<n>:LIMit<k>:CONTrol:MODE <Mode>
This command selects the horizontal limit line scaling.
Suffix:
<n>
.
Selects the measurement window.
<k> 1...8
Selects the limit line.
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values (Hz or s).
RELative
Limit line is defined by relative values related to the center fre-
quency (frequency domain) or the left diagram border (time
domain).
*RST: ABSolute
Example:
CALC:LIM2:CONT:MODE REL
Defines the x-axis of limit line 2 as relatively scaled.
CALCulate<n>:LIMit<k>:CONTrol:OFFSet <Offset>
This command defines an offset for a complete limit line.
Remote Control – Commands