Remote Control
R&S
®
FSVA/FSV
651Operating Manual 1176.7510.02 ─ 09
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Suffix:
<n>
.
Selects the measurement window.
<k> 1...8
Selects the limit line.
Parameters:
<Offset> Numeric value that defines the offset.
The unit of the offset depends on the scale of the x-axis.
*RST: 0
Example:
CALC:LIM2:CONT:OFFS 100us
Sets the X offset for limit line 2 (defined in zero span) to 100µs.
Manual operation: See "X Offset" on page 419
CALCulate<n>:LIMit<k>:CONTrol:SHIFt
<Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
Selects the measurement window.
<k> Selects the limit line.
Parameters:
<Distance> Numeric value that defines the distance of the shift.
The unit depends on the scale of the x-axis.
Example:
CALC:LIM2:CONT:SHIF 50KHZ
Shifts all reference values of limit line 2 by 50 kHz.
CALCulate<n>:LIMit<k>:CONTrol:SPACing <InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from frequency points.
Suffix:
<n>
.
Selects the measurement window.
<k> 1...8
Selects the limit line.
Parameters:
<InterpolMode> LINear | LOGarithmic
*RST: LIN
Example:
CALC:LIM:CONT:SPAC LIN
Remote Control – Commands