Remote Control
R&S
®
FSVA/FSV
684Operating Manual 1176.7510.02 ─ 09
<m> Selects the marker.
Return values:
<NumberOfPeaks>
Example:
CALC:MARK:FUNC:FPE 3
Searches the 3 highest maxima for trace 1
CALC:MARK:FUNC:FPE:COUN?
Queries the number of maxima found
Usage: Query only
Manual operation: See "Marker Peak List" on page 293
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE <MaxNoPeaks>
This command defines the maximum number of peaks the marker peak list may con-
tain.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<MaxNoPeaks> Maximum number of peaks to be determined.
*RST: 50
Example:
CALC:MARK:FUNC:FPE:LIST:SIZE 10
The marker peak list will contain a maximum of 10 peaks.
Manual operation: See "Max Peak Count" on page 293
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SEAR:AUTO <State>
This command turns the marker peak search on and off.
This command is retained for compatibility with R&S FSP only. Use CALCulate<n>:
MARKer<m>:FUNCtion:FPEaks:STAT instead.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:FPE:SEAR:AUTO ON
Activates marker peak search
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT <SortMode>
This command selects the order in which the results of a peak search are returned.