Remote Control
R&S
®
FSVA/FSV
685Operating Manual 1176.7510.02 ─ 09
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<SortMode> X
Sorts the peaks according to increasing position on the x-axis.
Y
Sorts the peaks according to decreasing position on the y-axis.
Example:
CALC:MARK:FUNC:FPE:SORT Y
Sets the sort mode to decreasing y values
Manual operation: See "Sort Mode Freq/Lvl" on page 293
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STAT <State>
This command turns a peak search on and off.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:FPE:STAT ON
Activates marker peak search
Manual operation: See "Peak List On/Off" on page 293
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X
This command queries the position of the peaks on the x-axis.
The order depends on the sort order that has been set with CALCulate<n>:
MARKer<m>:FUNCtion:FPEaks:SORT on page 684.
The number of peaks on the number that has been set with CALCulate<n>:
MARKer<m>:FUNCtion:FPEaks:COUNt?.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Return values:
<PeakPosition> Position of the peaks on the x-axis. The unit depends on the
measurement.
Remote Control – Commands

Table of Contents