804Operating Manual 1176.7510.02 ─ 09
Parameters:
<State> ON | OFF
*RST: ON
Example:
ESP:RANG3:SWE:TIME:AUTO OFF
Deactivates the sweep time auto mode for range 3.
Manual operation: See "Sweep Time Mode" on page 348
[SENSe:]ESPectrum:RANGe<range>:TRANsducer <TransducerName>
This command sets a transducer for the specified range. You can only choose a trans-
ducer that fulfills the following conditions:
The transducer overlaps or equals the span of the range.
The x-axis is linear.
The unit is dB.
Suffix:
<range>
.
1...20
range
Parameters:
<TransducerName> 'string' = name of the transducer
Example:
ESP:RANG1:TRAN 'test'
Sets the transducer called test for range 1.
Manual operation: See "Transd. Factor" on page 349
[SENSe:]ESPectrum:RRANge
This command returns the current position (number) of the reference range.
Example:
ESP:RRAN?
Returns the current position (number) of the reference range.
[SENSe:]ESPectrum:RTYPe <Type>
This command sets the power reference type.
Parameters:
<Type> PEAK | CPOWer
PEAK
Measures the highest peak within the reference range.
CPOWer
Measures the channel power within the reference range (integral
bandwidth method).
*RST: CPOWer
Example:
ESP:RTYP PEAK
Sets the peak power reference type.
Remote Control – Commands