Remote Control
R&S
®
FSVA/FSV
832Operating Manual 1176.7510.02 ─ 09
Return values:
<PulsePower> Lowest power level of the pulse power measurement.
The unit is dBm.
Example:
*ESE 1
*SRE 32
Configuration of status reporting systems for the generation of
an SRQ on operation complete
SENSe:MPOWer 935.2MHz,1MHz,434us,VIDEO,50PCT,
5us,MEAN,20;*OPC
Configuring and starting the measurement
Further actions of the control computer during measurement
On SRQ:
Response to service request
SENSe:MPOWer:RESult:MIN?
Usage:
Query only
[SENSe:]MPOWer[:SEQuence] <Frequency>, <RBW>, <MeasTime>,
<TriggerSource>, <TriggerLevel>, <TriggerOffset>, <Detector>, <#OfPulses>
This command configures and initiates the pulse power measurement.
The R&S FSVA/FSV caches all measurement parameters that you can set with this
command. If you use the command repeatedly, the R&S FSVA/FSV only changes
those settings that you have actually changed before initiating the measurement. Thus,
measurement times are kept as low as possible.
If you synchronize the measurement with *OPC, the R&S FSVA/FSV produces a ser-
vice request when all frequencies have been measured and the number of individual
measurements has been performed.
Note that using the command as a query initiates the measurement and returns the
results if all frequencies have been measured. For more information on querying the
results see .
To reduce the setting time, the setup is performed simultaneously for all selected
parameters.
The query adapts the instrument settings, performs the defined number of measure-
ments and outputs the measurement results list.
This measurement is not compatible with other measurements, especially as far as
marker functions, adjacent-channel measurement or statistics are concerned. The cor-
responding functions are therefore automatically switched off. In return, incompatible
commands automatically deactivate the multi burst power function.
The function is only available in the REMOTE operation. It is deactivated on switching
back to LOCAL.
Parameters:
<Frequency> Defines the pulse frequency.
Range: 0 Hz to fmax
Remote Control – Commands