Remote Control
R&S
®
FSVA/FSV
831Operating Manual 1176.7510.02 ─ 09
[SENSe:]MPOWer:FTYPe...............................................................................................831
[SENSe:]MPOWer:RESult[:LIST]?................................................................................... 831
[SENSe:]MPOWer:RESult:MIN?......................................................................................831
[SENSe:]MPOWer[:SEQuence]....................................................................................... 832
[SENSe:]MPOWer:FTYPe <FilterType>
This command selects the filter type for pulse power measurements.
Parameters:
<FilterType> NORMal | CFILter | RRC
Example:
SENSe:MPOWer:FTYPe CFILter
[SENSe:]MPOWer:RESult[:LIST]?
This command queries the results of the pulse power measurement.
This command may be used to obtain measurement results in an asynchronous way,
using the service request mechanism for synchronization to the end of the measure-
ment.
If there are no results, the command returns an error.
Parameters:
<PulsePower> List of pulse powers.
The number of values depends on the number of pulses you
have been measuring.
The unit is dBm.
Example:
*ESE 1
*SRE 32
Configuration of status reporting systems for the generation of
an SRQ on operation complete
SENSe:MPOWer 935.2MHz,1MHz,434us,VIDEO,50PCT,
5us,MEAN,20;*OPC
Configuring and starting the measurement
…
Further actions of the control computer during measurement
On SRQ:
Response to service request
SENSe:MPOWer:RESult?
Usage:
Query only
[SENSe:]MPOWer:RESult:MIN?
This command queries the lowest pulse power that has been measured during a pulse
power measurement.
If there are no results, the command returns an error.
Remote Control – Commands