Remote Control
R&S
®
FSVA/FSV
846Operating Manual 1176.7510.02 ─ 09
For the relative measurement the reference value is set to the currently measured
channel power using the command [SENSe:]POWer:ACHannel:REFerence:
AUTO ONCE.
Parameters:
<Mode> ABSolute | RELative
ABSolute
absolute adjacent channel measurement
RELative
relative adjacent channel measurement
*RST: RELative
Example:
POW:ACH:MODE REL
Sets the adjacent channel measurement mode to relative.
Manual operation: See "ACLR (Abs/Rel)" on page 331
[SENSe:]POWer:ACHannel:NAME:ACHannel <Name>
This command defines a name for the adjacent channel. The name is displayed in the
result diagram and the result table.
Parameters:
<Name> *RST: Adj
Example:
POW:ACH:NAME:ACH 'XYZ'
Defines the name of the adjacent channel as 'XYZ'.
Manual operation: See "Names" on page 329
[SENSe:]POWer:ACHannel:NAME:ALTernate<channel> <Name>
This command defines a name for the specified alternate channel. The name is dis-
played in the result diagram and the result table.
Suffix:
<channel>
.
1...11
the alternate channel
Parameters:
<Name> *RST: ALT<1...11>
Example:
POW:ACH:NAME:ALT3 'XYZ'
Defines the name of the third alternate channel as 'XYZ'.
Manual operation: See "Names" on page 329
[SENSe:]POWer:ACHannel:NAME:CHANnel<channel> <Name>
This command defines a name for the specified TX channel. The name is displayed in
the result diagram and the result table.
Remote Control – Commands