Instrument Functions
R&S
®
FSVA/FSV
329Operating Manual 1176.7510.02 ─ 09
Names ← Channel Setup ← CP/ACLR Settings ← Ch Power ACLR
Define user-specific channel names for each channel. The names defined here are dis-
played in the result diagram and result table.
Remote command:
[SENSe:]POWer:ACHannel:NAME:ACHannel on page 846
[SENSe:]POWer:ACHannel:NAME:ALTernate<channel> on page 846
[SENSe:]POWer:ACHannel:NAME:CHANnel<channel> on page 846
Weighting Filter ← Channel Setup ← CP/ACLR Settings ← Ch Power ACLR
Define weighting filters for all channels. Weighting filters are not available for all sup-
ported standards and cannot always be defined manually where they are available.
The dialog contains the following fields:
Field Description
Channel
TX 1-18: TX channels
ADJ: Adjacent channel
ALT1-11: Alternate channels
Active Activates/Deactivates the weighting filter for the selected and any subsequent channels of
the same type
Alpha Defines the alpha value for the weighting filter for the selected and any subsequent chan-
nels of the same type
Remote command:
POW:ACH:FILT:CHAN1 ON, see [SENSe:]POWer:ACHannel:FILTer[:STATe]:
CHANnel<channel> on page 845
Activates the weighting filter for TX channel 1.
POW:ACH:FILT:ALPH:CHAN1 0,35 see [SENSe:]POWer:ACHannel:FILTer:
ALPHa:CHANnel<channel> on page 844
Sets the alpha value for the weighting filter for TX channel 1 to 0,35.
POW:ACH:FILT:ACH ON see [SENSe:]POWer:ACHannel:FILTer[:STATe]:
ACHannel on page 845
Activates the weighting filter for the adjacent channel.
POW:ACH:FILT:ALPH:ACH 0,35 see [SENSe:]POWer:ACHannel:FILTer:
ALPHa:ACHannel on page 843
Sets the alpha value for the weighting filter for the adjacent channel to 0,35.
POW:ACH:FILT:ALT1 ON see [SENSe:]POWer:ACHannel:FILTer[:STATe]:
ALTernate<channel> on page 845
Activates the alpha value for the weighting filter for the alternate channel 1.
POW:ACH:FILT:ALPH:ALT1 0,35 see [SENSe:]POWer:ACHannel:FILTer:
ALPHa:ALTernate<channel> on page 844
Sets the alpha value for the weighting filter for the alternate channel 1 to 0,35.
Limits ← Channel Setup ← CP/ACLR Settings ← Ch Power ACLR
Activate and define the limits for the ACLR measurement.
Limit Checking ← Limits ← Channel Setup ← CP/ACLR Settings ← Ch Power
ACLR
Activate or deactivate limit checking for the ACLR measurement.
Measurement Functions