Remote Control
R&S
®
FSVA/FSV
843Operating Manual 1176.7510.02 ─ 09
With [SENSe<source>:]POWer:HSPeed set to ON, steep-edged channel filters are
available. For further information on filters refer to Chapter 3.2.6.4, "List of Available
RRC and Channel Filters", on page 244 .
Parameters:
<Bandwidth> 100 Hz to 40 GHz
*RST: 14 kHz
Example:
POW:ACH:BWID:ACH 30 kHz
Sets the bandwidth of all adjacent channels to 30 kHz.
Manual operation: See "Bandwidth" on page 327
[SENSe:]POWer:ACHannel:BANDwidth|BWIDth:ALTernate<channel>
<Bandwidth>
This command defines the channel bandwidth of the specified alternate adjacent chan-
nels of the radio transmission system. If the channel bandwidth of one alternate adja-
cent channel is changed (e.g. channel 3), the bandwidth of all subsequent alternate
adjacent channels (e.g. 4–11) is automatically set to the same value.
With [SENSe<source>:]POWer:HSPeed set to ON, steep-edged channel filters are
available. For further information on filters refer to Chapter 3.2.6.4, "List of Available
RRC and Channel Filters", on page 244 .
Suffix:
<channel>
.
1...11
the alternate adjacent channel
Parameters:
<Bandwidth> 100 Hz to 40 GHz
*RST: 14 kHz
Example:
POW:ACH:BWID:ALT2 30 kHz
Manual operation: See "Bandwidth" on page 327
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel <Alpha>
This command defines the roll-off factor for the weighting filter of the adjacent channel.
Parameters:
<Alpha> <numeric value>
*RST: 0,22
Example:
POW:ACH:FILT:ALPH:ACH 0,35
Manual operation: See "Weighting Filter" on page 329
[SENSe:]POWer:ACHannel:FILTer:ALPHa[:ALL] <Alpha>
This command defines the roll-off factor for the weighting filter of all channels.
Remote Control – Commands