Remote Control
R&S
®
FSVA/FSV
844Operating Manual 1176.7510.02 ─ 09
Parameters:
<Alpha> <numeric value>
*RST: 0,22
Example:
POW:ACH:FILT:ALPH:ALL 0,35
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<channel> <Alpha>
This command defines the roll-off factor for the weighting filter of the specified alternate
channel.
Suffix:
<channel>
.
1...11
the alternate channel
Parameters:
<Alpha> <numeric value>
*RST: 0,22
Example:
POW:ACH:FILT:ALPH:ALT3 0,35
Sets the alpha value for the weighting filter for the alternate
channel 3 to 0,35.
Manual operation: See "Weighting Filter" on page 329
[SENSe:]POWer:ACHannel:FILTer:ALPHa:CHANnel<channel> <Alpha>
This command defines the roll-off factor for the weighting filter of the specified TX
channel.
Suffix:
<channel>
.
1...11
the TX channel
Parameters:
<Alpha> <numeric value>
*RST: 0,22
Example:
POW:ACH:FILT:ALPH:CHAN3 0,35
Sets the alpha value for the weighting filter for the TX channel 3
to 0,35.
Manual operation: See "Weighting Filter" on page 329
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALL <State>
This command activates the weighting filter for all channels.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
POW:ACH:FILT ON
Remote Control – Commands