Remote Control
R&S
®
FSVA/FSV
851Operating Manual 1176.7510.02 ─ 09
[SENSe:]POWer:HSPeed <State>
This command switches on or off the high-speed channel/adjacent channel power
measurement. The measurement itself is performed in zero span on the center fre-
quencies of the individual channels. The command automatically switches to zero span
and back.
Depending on the selected mobile radio standard, weighting filters with characteristic
or very steep-sided channel filters are used for band limitation.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
POW:HSP ON
Manual operation: See "Fast ACLR (On/Off)" on page 332
[SENSe:]POWer:NCORrection
<Mode>
This command turns noise cancellation on and off.
If noise cancellation is on, the R&S FSVA/FSV performs a reference measurement to
determine its inherent noise and subtracts the result from the channel power measure-
ment result (first active trace only).
The inherent noise of the instrument depends on the selected center frequency, resolu-
tion bandwidth and level setting. Therefore, the correction function is disabled when-
ever one of these parameters is changed. A corresponding message is displayed on
the screen. Noise correction must be turned on again manually after the change.
Parameters:
<Mode> ON
Performs noise correction.
OFF
Performs no noise correction.
AUTO
Performs noise correction.
After a parameter change, noise correction is restarted automati-
cally and a new correction measurement is performed.
*RST: OFF
Example:
POW:NCOR ON
Manual operation: See "Noise Correction" on page 231
[SENSe:]POWer:TRACe <TraceNumber>
This command assigns the channel/adjacent channel power measurement to the indi-
cated trace. The corresponding trace must be active, i.e. its state must be different
from blank.
Remote Control – Commands