Remote Control
R&S
®
FSVA/FSV
859Operating Manual 1176.7510.02 ─ 09
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Manual operation: See "Sweeptime Manual" on page 239
See "Sweeptime Auto" on page 240
See "Default Coupling" on page 242
[SENSe:]SWEep:TYPE <Type>
This command selects the sweep type.
Parameters:
<Type> SWE
Selects analog frequency sweeps.
AUTO
Automatically selects the sweep type (FFT or analog frequency
sweep).
FFT
Selects FFT sweeps.
*RST: AUTO
Example:
SWE:TYPE FFT
Selects FFT sweeps.
Manual operation: See "Sweep" on page 236
See "FFT" on page 236
See "Auto" on page 236
[SENSe:]SWEep:TYPE:USED?
This command queries the sweep type if you have turned on automatic selection of the
sweep type.
Return values:
<Type> SWE | FFT
SWE
Normal sweep
FFT
FFT mode
*RST: SWE
Example:
SWE:TYPE:USED?
Usage: Query only