Remote Control
R&S
®
FSVA/FSV
858Operating Manual 1176.7510.02 ─ 09
Example:
SWE:MODE ESP
Sets the Spectrum Emission Mask measurement mode.
Manual operation: See "Spectrum Emission Mask" on page 312
See "Spurious Emissions" on page 312
See "Spectrum Emission Mask" on page 346
See "Spurious Emissions" on page 378
[SENSe:]SWEep:POINts <NumberPoints>
This command defines the number of measurement points to be collected during one
sweep.
Note: For Spurious Emissions measurements the maximum number of sweep points in
all ranges is limited to 100001.
Parameters:
<NumberPoints> Range: 101 to 32001
*RST: 691
Example:
SWE:POIN 251
Manual operation: See "Sweep Points" on page 249
[SENSe:]SWEep:TIME <Time>
This command defines the sweep time.
The range depends on the frequency span.
If this command is used in "Spectrum" mode, automatic coupling to resolution band-
width and video bandwidth is switched off.
Parameters:
<Time> refer to data sheet
*RST: (automatic)
Example:
SWE:TIME 1s
Manual operation: See "Sweeptime Manual" on page 239
See "Sweep Time" on page 332
See "Meas Time" on page 447
[SENSe:]SWEep:TIME:AUTO <State>
In realtime mode, this command automatically sets the sweep time to 32 ms.
In analyzer mode, this command controls the automatic coupling of the sweep time to
the frequency span and bandwidth settings. If [SENSe:]SWEep:TIME is used, auto-
matic coupling is switched off.
Parameters:
<State> ON | OFF
*RST: ON
Remote Control – Commands