Remote Control
R&S
®
FSVA/FSV
872Operating Manual 1176.7510.02 ─ 09
STATus:QUEStionable:ACPLimit[:EVENt]?
This command queries the contents of the "EVENt" part of the
STATus:QUEStionable:ACPLimit register. Readout deletes the contents of the
"EVENt" part.
Example:
STAT:QUES:ACPL?
Usage: Query only
Mode: all
STATus:QUEStionable:ACPLimit:CONDition?
This command queries the contents of the "CONDition" part of the
STATus:QUEStionable:ACPLimit register (see "STATus:QUEStionable:ACPLimit
Register" on page 594). Readout does not delete the contents of the "CONDition" part.
Example:
STAT:QUES:ACPL:COND?
Usage: Query only
Mode: all
STATus:QUEStionable:ACPLimit:ENABle <BitDefinition>
This command sets the bits of the "ENABle" part of the
STATus:QUEStionable:ACPLimit register. The "ENABle" register selectively ena-
bles the individual events of the associated "EVENt" part for the summary bit.
Parameters:
<BitDefinition> 0 to 65535
Example:
STAT:QUES:ACPL:ENAB 65535
Mode: all
STATus:QUEStionable:ACPLimit:NTRansition <BitDefinition>
This command sets the edge detectors of all bits of the
STATus:QUEStionable:ACPLimit register from 1 to 0 for the transitions of the
"CONDition" bit.
Parameters:
<BitDefinition> 0 to 65535
Example:
STAT:QUES:ACPL:NTR 65535
Mode: all