873Operating Manual 1176.7510.02 ─ 09
STATus:QUEStionable:ACPLimit:PTRansition <BitDefinition>
This command sets the edge detectors of all bits of the
STATus:QUEStionable:ACPLimit register from 0 to 1 for the transitions of the
"CONDition" bit.
Parameters:
<BitDefinition> 0 to 65535
Example:
STAT:QUES:ACPL:PTR 65535
Mode: all
STATus:QUEStionable:DIQ:CONDition?
This command queries the contents of the "CONDition" part of the
STATus:QUEStionable:DIQ register for IQ measurements (see "STATus:QUES-
tionable:DIQ Register" on page 595). Readout does not delete the contents of the
"CONDition" part.
Example:
STAT:QUES:DIQ:COND?
Usage: Query only
Mode: IQ
STATus:QUEStionable:DIQ[:EVENt]?
This command queries the contents of the "EVENt" part of the
STATus:QUEStionable:DIQ register for IQ measurements.
Readout deletes the contents of the "EVENt" part.
Example:
STAT:QUES:DIQ?
Usage: Query only
Mode: IQ
STATus:QUEStionable:DIQ:ENABle <Enable>
Setting parameters:
<Enable> numeric value
Usage: SCPI confirmed
Mode: all
STATus:QUEStionable:DIQ:NTRansition <NTransition>
Setting parameters:
<NTransition> numeric value
Usage: SCPI confirmed