Remote Control
R&S
®
FSVA/FSV
893Operating Manual 1176.7510.02 ─ 09
Parameters:
<NumberSets> 0 to 32767
*RST: 0
Example:
TRAC:IQ ON
Switches on acquisition of I/Q data.
TRAC:IQ:AVER ON
Enables averaging of the I/Q measurement data
TRAC:IQ:AVER:COUN 10
Selects averaging over 10 data sets
TRAC:IQ:DATA?
Starts the measurement and reads out the averaged data.
Usage: SCPI confirmed
TRACe<n>:IQ:AVERage[:STATe]
<State>
The command enables averaging of the recorded I/Q data, provided that I/Q data
acquisition was previously enabled with TRACe<n>:IQ[:STATe] on page 892.
With I/Q data averaging enabled, the maximum amount of I/Q data is limited to 512kS
(524288 complex samples).
Suffix:
<n>
.
irrelevant
Parameters:
<State> ON | OFF
*RST: OFF
Example:
TRAC:IQ ON
Switches on acquisition of I/Q data.
TRAC:IQ:AVER ON
Enables averaging of the I/Q measurement data.
TRAC:IQ:AVER:COUN 10
Selects averaging over 10 data sets.
TRAC:IQ:DATA?
Starts the measurement and reads out the averaged data.
TRACe<n>:IQ:BWIDth
This command queries the flat, usable filter bandwidth of the final I/Q data.
The bandwidth depends on the defined user sample rate and the defined signal source
(RF/Digital IQ), see also Figure 3-37. This value can not be changed by the user.
Suffix:
<n>
.
irrelevant
Example:
TRAC:IQ:BWID?
Manual operation: See "Filter BW" on page 446
See "Maximum Bandwidth" on page 446
Remote Control – Commands