Remote Control
R&S
®
FSVA/FSV
942Operating Manual 1176.7510.02 ─ 09
MMEMory:LOAD:AUTO <Mode>
This command defines which settings file is automatically loaded after the device is
switched on. The contents of the file are read after switching on the device and used to
define the new device state.
The settings file defined as auto recall set can also be restored by the *RST
on page 611 command.
Parameters:
<Mode> 1,'FACTORY' | 1,'<file_name>' | with <file_name>
1,'FACTORY'
deactivates the startup recall function
1,'<file_name>'
activates the startup recall function and defines the settings file
to be loaded
with <file_name>
DOS file name with or without extension
The file name includes indication of the path and may also
include the drive name. The path name complies with DOS con-
ventions.
Example:
MMEM:LOAD:AUTO 1,'C:\R_S\Instr\user\TEST'
Usage: Event
Manual operation: See "Startup Recall (On/Off)" on page 200
MMEMory:LOAD:STATe 1, <FileName>
This command restores and activates an instrument configuration stored in a *.dfl file.
Note that files with other formats cannot be loaded with this command.
The contents that are reloaded from the file are defined by the last selection made
either in the "Save/Recall" dialogs (manual operation) or through the
MMEMory:SELect[:ITEM] commands (remote operation; the settings are identical in
both cases).
By default, the selection is limited to the user settings ("User Settings" selection in the
dialogs, HWSettings in SCPI). The selection is not reset by PRESET or *RST.
As a consequence, the results of a SCPI script using the MMEMory:LOAD:STATe com-
mand without a previous MMEMory:SELect[:ITEM] command may vary, depending
on previous actions in the GUI or in previous scripts, even if the script starts with the
*RST command.
It is therefore recommended that you use the appropriate MMEMory:SELect[:ITEM]
command before using MMEMory:LOAD:STATe.
Parameters:
1