Instrument Function
R&S
®
SMB100A
149Operating Manual 1407.0806.32 ─ 13
RF Frequency Sweep Settings
State - Frequency Sweep
Activates/deactivates RF sweep mode.
Note:
Activating a sweep mode automatically deactivates other sweeps and the list mode.
Remote command:
[:SOURce<hw>]:FREQuency:MODE on page 306
Mode - Frequency Sweep
Selects the sweep instrument operating mode and the sweep mode.
If you change the sweep mode during the execution, the signal generator stops the sweep
and starts with the next trigger event at the initial value.
The "Reset Sweep" button sets the sweep to the start value.
"Auto"
Sets an automatic repeated sweep cycle.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR AUTO
SOUR:FREQ:MODE SWE
"Single"
Sets a single sweep cycle. The sweep is triggered by the "Execute Sin-
gle Sweep" button, or by means remote trigger commands, e.g. *TRG.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR SING
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:EXEC